The Hahn™ Tapered Implant Guided Mount is designed for the manual seating and rotation of a Hahn Tapered Implant. The Guided Mount is used in conjunction with the Ratchet Wrench and a Surgical Adaptor during guided dental implant placement. If needed, the Hahn Tapered Implant Mount Wrench may be used to rotate the Guided Mount to achieve proper alignment between the mount and its corresponding guide sleeve.
Each Hahn Tapered Implant Guided Mount comes packaged with a captured screw.
Patients should be evaluated before the time of surgery for factors that put them at risk from the implant placement procedure, or that may affect healing of bone or surrounding soft tissue. Contraindications include but are not limited to: vascular conditions, uncontrolled diabetes, clotting disorders, anticoagulant therapy, metabolic bone disease, chemotherapy or radiation therapy, chronic periodontal inflammation, insufficient soft tissue coverage, metabolic or systemic disorders associated with wound and/or bone healing, use of pharmaceuticals that inhibit or alter natural bone remodeling, any disorders which inhibit a patient's ability to maintain adequate daily oral hygiene, uncontrolled parafunctional habits, insufficient height and/or width of bone, and insufficient interarch space.
Do NOT rotate after the flange on the Guided Mount fully meets the guide sleeve and the corresponding hexes are aligned. Doing so may cause the osteotomy to strip.
Hahn Tapered Implant Guided Mounts are shipped NON-STERILE. Non-sterile items must be cleaned, disinfected, and sterilized prior to clinical use, according to a validated method.
The recommended disinfection process is based on ANSI/AAMI ST79 guidelines, as follows:
The recommended sterilization process is based on the ANSI/AAMI/ISO 17665-1 and ANSI/AAMI ST79 guidelines, as follows:
NOTE: The validated procedures require the use of FDA-cleared sterilization trays, wraps, biological indicators, chemical indicators, and other sterilization accessories labeled for the sterilization cycle recommended. The healthcare facility should monitor the sterilizer for the facility according to an FDA-recognized sterility assurance standard such as ANSI/AAMI ST79.
1Oral disinfectant containing Chlorhexidine is recommended; refer to the disinfectant manufacturer's instructions.
Diameter: 4.3 mm
Connection Type: Unigrip screw / Ø mm conical male / 4 mm male hex